STATus:OPERation[:EVENt]?
Returns the sum of the bits in the event register for the Standard Operation Register group. An event
register is a read-only register that latches events from the condition register. While an event bit is set, sub-
sequent events corresponding to that bit are ignored. The register bits are cleared when you read the
register.
Parameter Typical Return
(none) +512
Read the event register (bit 9 is set):
STAT:OPER:EVEN?
l Once a bit is set, it remains set until cleared by reading the event register or by sending *CLS (clear
status).
STATus:PRESet
Clears Questionable Data enable register and Standard Operation enable register.
Parameter Typical Return
(none) (none)
Clear enable register bits:
STAT:PRES
STATus:QUEStionable:CONDition?
Returns the sum of the bits in the condition register for the Questionable Data Register group. This register
is read-only; bits are not cleared when read.
A condition register continuously monitors the state of the instrument. Condition register bits are updated
in real time; they are neither latched nor buffered.
Parameter Typical Return
(none) +4096
Read the condition register (bit 12 is set):
STAT:QUES:COND?
l The condition register bits reflect the current condition. If a condition goes away, the corresponding
bit is cleared in the condition register.
SCPI Programming Reference
390
Keysight Truevolt Series Operating and Service Guide