Math Operations Factory Setting
Math State Disabled
Scale Function Null
Null and dB Relative Value Set to 0.0 with auto reference enabled
Upper and Lower Limits 0.0
Histogram Bin counts cleared; 100 bins; auto binning enabled
Statistics Cleared
• dBm Reference Resistance • 600 Ω
Calculate Scale (34465A/70A only):
Gain 1.0
Offset 0.0
Reference 0.0
Unit State Off
Unit Null String
Calculate Smoothing (34465A/70A only):
Response Fast
State Off
Triggering Operations Factory Setting
Trigger Count 1
Trigger Source Immediate
Trigger Delay Auto Delay
Sample Count 1
Trigger Slope Negative
VM Comp Output Slope NEG
Pretrigger Count (34465A/70A only) 0
Triggering (34465A/70A only):
Sample Source Immediate
Sample Timer 1 s
Trigger Level 0.0
System-Related Operations Factory Setting
• Beeper Mode • On
• Thousands Separator • On
Display State On
Reading Memory Cleared
Error Queue See note
• Stored States • No Change
• Calibration State • Secured
Keysight Truevolt Series Operating and Service Guide
475
SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals