Values for 34460A DMM
Input Error from Nominal
Resistance Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
100
4-wire Ohms
100 ± 10 mΩ ± 18 m ± 21 m ± 24 m
1 kΩ 1 kΩ Q ± 38 mΩ ± 120 m ± 150 m ± 180 m
10 kΩ 10 kΩ ± 350 m ± 1.2 ± 1.5 ± 1.8
100 kΩ 100 kΩ ± 3.5 ± 12 ± 15 ± 18
1 MΩ 1 MΩ ± 40 ± 120 ± 150 ± 180
10 MΩ 10 MΩ Q ± 1.6 ± 2.1 ± 4.1 ± 6.1
100 MΩ 2-wire Ohms 100 MΩ ± 310 ± 810
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Values for 34461A DMM
Input Error from Nominal
Voltage Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
100 mV
DC Volts
100 mV ± 6 µV ± 7.5 µV ± 8.5 µV ± 10 µV
-100 mV
1 V
1 V ± 26 µV ± 37 µV ± 47 µV ± 62 µV
-1 V
4 V
10 V
± 100 µV ± 130 µV ± 190 µV ± 230 µV
10 V Q
± 190 µV ± 250 µV ± 400 µV ± 500 µV
-10 V
100 V
100 V ± 2.6 mV ± 4.1 mV ± 5.1 mV ± 6.6 mV
-100 V Q
1000 V
1000 V
± 26 mV ± 45 mV ± 55 mV ± 70 mV
-500 V ± 16 mV ± 27.5 mV ± 32.5 mV ± 40 mV
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Keysight Truevolt Series Operating and Service Guide
511
Performance Verification