Values for 34461A DMM
Input Error from Nominal
Current Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
2 A
DC Current
3 A ± 4.2 mA ± 4.6 mA ± 4.6 mA ± 5.2 mA
1 A 1 A Q ± 560 µA ± 900 µA ± 1.1 mA ± 1.3 mA
100 mA 100 mA ± 14 µA ± 35 µA ± 55 µA ± 65 µA
10 mA 10 mA Q ± 2.7 µA ± 5 µA ± 7 µA ± 8 µA
1 mA 1 mA ± 0.13 µA ± 0.36 µA ± 0.56 µA ± 0.66 µA
100 µA 100 µA ± 0.03 µA ± 0.065 µA ± 0.075 µA ± 0.085 µA
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Values for 34461A DMM
Input Error from Nominal
Resistance Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
100 Ω
4-wire Ohms
100 Ω ± 6 mΩ ± 12 mΩ ± 14 mΩ ± 16 mΩ
1 kΩ 1 kΩ Q ± 25 mΩ ± 90 mΩ ± 110 mΩ ± 130 mΩ
10 kΩ 10 kΩ ± 250 mΩ ± 900 mΩ ± 1.1 Ω ± 1.3 Ω
100 kΩ 100 kΩ ± 2.5 Ω ± 9 Ω ± 11 Ω ± 13 Ω
1 MΩ 1 MΩ ± 30 Ω ± 90 Ω ± 110 Ω ± 130 Ω
10 MΩ 10 MΩ Q ± 1.6 kΩ ± 2.1 kΩ ± 4.1 kΩ ± 6.1 kΩ
100 MΩ 2-wire Ohms 100 MΩ ± 310 kΩ ± 810 kΩ
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Values for 34465A DMM
Input Error from Nominal
Current Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
2 A
DC Current
3 A ± 4.2 mA ± 4.6 mA ± 4.6 mA ± 5.2 mA
1 A 1 A Q ± 560 µA ± 800 µA ± 900 µA ± 1.1 mA
100 mA 100 mA ± 14 µA ± 35 µA ± 55 µA ± 65 µA
10 mA 10 mA Q ± 2.7 µA ± 5 µA ± 7 µA ± 8 µA
1 mA 1 mA ± 0.1 µA ± 0.35 µA ± 0.55 µA ± 0.65 µA
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Performance Verification
512