Values for 34465A DMM
Input Error from Nominal
Voltage Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
100 mV
DC Volts
100 mV ± 6 µV ± 7.5 µV ± 8.5 µV ± 10 µV
-100 mV
1 V
1 V ± 19 µV ± 29 µV ± 39 µV ± 54 µV
-1 V
4 V
10 V
± 70 µV ± 120 µV ± 160 µV ± 220 µV
10 V Q
± 130 µV ± 240 µV ± 340 µV ± 490 µV
-10 V
100 V
100 V ± 2.6 mV ± 4.1 mV ± 4.6 mV ± 6.1 mV
-100 V Q
1000 V
1000 V
± 36 mV ± 51 mV ± 56 mV ± 71 mV
-500 V ± 16 mV ± 23.5 mV ± 26 mV ± 33.5 mV
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Values for 34465A DMM
Input Error from Nominal
Resistance Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
100
4-wire Ohms
100 ± 6 mΩ ± 9 m ± 10 m ± 11 m
1 kΩ 1 kΩ Q ± 25 mΩ ± 35 mΩ ± 45 m ± 55 m
10 kΩ 10 kΩ ± 250 m ± 350 m ± 450 m ± 550 m
100 kΩ 100 kΩ ± 2.5 ± 3.5 ± 4.5 ± 5.5
1 MΩ 1 MΩ ± 25 ± 65 ± 75 ± 85
10 MΩ 10 MΩ
± 1.1 ± 2.1 ± 2.6 ± 3.1
10 MΩ
2-wire Ohms
10 MΩ Q
100 MΩ 100 MΩ ± 101 ± 201 ± 301 ± 401
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Keysight Truevolt Series Operating and Service Guide