Values for 34470A DMM
Input Error from Nominal
Current Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
2 A
DC Current
3 A ± 4.2 mA ± 4.6 mA ± 5.2 mA
1 A 1 A Q ± 560 µA ± 800 µA ± 900 µA ± 1.1 mA
100 mA 100 mA ± 14 µA ± 35 µA ± 55 µA ± 65 µA
10 mA 10 mA Q ± 2.7 µA ± 5 µA ± 7 µA ± 8 µA
1 mA 1 mA ± 0.1 µA ± 0.35 µA ± 0.55 µA ± 0.65 µA
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Values for 34470A DMM
Input Error from Nominal
Voltage Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
100 mV
DC Volts
100 mV ± 6 µV ± 7.5 µV ± 8 µV
-100 mV
1 V
1 V ± 14 µV ± 19 µV ± 24 µV ± 29 µV
-1 V
4 V
10 V
± 52 µV ± 72 µV ± 84 µV ± 100 µV
10 V Q
± 100 µV ± 150 µV ± 180 µV ± 220 µV
-10 V
100 V
100 V ± 2.6 mV ± 3.8 mV ± 4.4 mV ± 4.6 mV
-100 V Q
1000 V
1000 V
± 36 mV ± 48 mV ± 54 mV ± 56 mV
-500 V ± 16 mV ± 22 mV ± 25 mV ± 36 mV
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Values for 34470A DMM
Input Error from Nominal
Resistance Function
[1]
Range
Quick
Check
24 hours 90 days 1 year 2 years
100
4-wire Ohms
100 ± 6 mΩ ± 9 m ± 10 m ± 11 m
1 kΩ 1 kΩ Q ± 25 mΩ ± 35 mΩ ± 45 m ± 55 m
10 kΩ 10 kΩ ± 250 m ± 350 m ± 450 m ± 550 m
100 kΩ 100 kΩ ± 2.5 ± 3.5 ± 4.5 ± 5.5
1 MΩ 1 MΩ ± 25 ± 65 ± 75 ± 85
10 MΩ 10 MΩ
± 1.1 ± 2.1 ± 2.6 ± 3.1
10 MΩ
2-wire Ohms
10 MΩ Q
100 MΩ 100 MΩ ± 101 ± 201 ± 301 ± 401
[1]
Select 10 PLC.
Q: Quick performance verification test point.
Performance Verification