ACV Values for 34461A DMM
Input Error from Nominal
Vrms Frequency Range Filter
Quick
Check
24 hours 90 days 1 year 2 years
100 mV
1 kHz
100 mV 200 Hz
± 60 µV ± 80 µV ± 90 µV ± 100 µV
50 kHz ± 140 µV ± 160 µV ± 170 µV ± 180 µV
300 kHz Q ± 4.5 mV
1 V
1 kHz
1 V 200 Hz
± 600 µV ± 800 µV ± 900 µV ± 1 mV
50 kHz ± 1.4 mV ± 1.6 mV ± 1.7 mV ± 1.8 mV
300 kHz ± 45 mV
0.03 V 1 kHz
10 V
200 Hz ± 2 mV ± 3 mV
1 V 1 kHz 200 Hz ± 2.4 mV ± 3.5 mV ± 3.6 mV ± 3.7 mV
10 V
10 Hz 3 Hz ± 6 mV ± 8 mV ± 9 mV ± 10 mV
100 Hz 20 Hz ± 6 mV ± 8 mV ± 9 mV ± 10 mV
20 kHz 200 Hz ± 6 mV ± 8 mV ± 9 mV ± 10 mV
50 kHz
200 Hz
Q ± 14 mV ± 16 mV ± 17 mV ± 18 mV
100 kHz ± 63 mV ± 68 mV
300 kHz ± 450 mV
100 V
1 kHz
100 V 200 Hz
Q ± 60 mV ± 80 mV ± 90 mV ± 100 mV
50 kHz ± 140 mV ± 160 mV ± 170 mV ± 180 mV
70 V 300 kHz ± 3.3 V
750 V 1 kHz
750 V 200 Hz
± 450 mV ± 600 mV ± 675 mV ± 750 mV
210 V 50 kHz ± 510 mV ± 606 mV ± 627 mV ± 648 mV
70 V 300 kHz ± 6.6 V
Q: Quick performance verification test point.
Keysight Truevolt Series Operating and Service Guide
517
Performance Verification