ACI Values for 34460A and 34461A DMMs
Input Error from Nominal
Current, rms Frequency Range Filter
Quick
Check
24 hours 90 days 1 year 2 years
2 A
1 kHz
3 A 200 Hz ± 5.8 mA
5 kHz
1 A
1 kHz
1 A 200 Hz ± 1.4 mA
5 kHz
100 mA
10 Hz
100 mA
3 Hz
± 140 µA1 kHz 200 Hz
5 kHz 200 Hz
100 µA 1 kHz
10 mA 200 Hz
4.1 µA
1 mA 1 kHz 5 µA
10 mA
1 kHz
14 µA
5 kHz
1 mA
1 kHz
1 mA 200 Hz 1.4 µA
5 kHz
100 µA
1 kHz
100 µA 200 Hz 0.14 µA
5 kHz Q
Q: Quick performance verification test point.