ACV Values for 34465A and 34470A DMMs
Input Error from Nominal
Vrms Frequency Range Filter
Quick
Check
24 hours 90 days 1 year 2 years
100 mV
1 kHz
100 mV 200 Hz
± 40 µV ± 60 µV ± 70 µV ± 80 µV
50 kHz ± 80 µV ± 90 µV ± 100 µV ± 110 µV
300 kHz Q ± 1.1 mV
1 V
1 kHz
1 V 200 Hz
± 400 µV ± 600 µV ± 700 µV ± 800 µV
50 kHz ± 800 µV ± 900 µV ± 1 mV ± 1.1 mV
300 kHz ± 11 mV
0.03 V 1 kHz
10 V
200 Hz ± 2 mV
1 V 1 kHz 200 Hz ± 2.2 mV ± 2.4 mV ± 2.5 mV ± 2.6 mV
10 V
10 Hz 3 Hz
± 4 mV ± 6 mV ± 7 mV ± 8 mV100 Hz 20 Hz
20 kHz 200 Hz
50 kHz
200 Hz
Q ± 8 mV ± 9 mV ± 10 mV ± 11 mV
100 kHz ± 20 mV
300 kHz ± 110 mV
100 V
1 kHz
100 V 200 Hz
Q ± 40 mV ± 60 mV ± 70 mV ± 80 mV
50 kHz ± 80 mV ± 90 mV ± 100 mV ± 110 mV
70 V 300 kHz ± 800 mV
750 V 1 kHz
750 V 200 Hz
± 750 mV ± 900 mV ± 975 mV ± 1.1 V
210 V 50 kHz ± 330 mV ± 351 mV ± 372 mV ± 393 mV
70 V 300 kHz ± 1.5 V
Q: Quick performance verification test point.
ACI Values for 34465A and 34470A DMMs
Input Error from Nominal
Current, rms Frequency Range Filter
Quick
Check
24 hours 90 days 1 year 2 years
2 A
1 kHz
3 A 200 Hz ± 5.8 mA
5 kHz
1 A
1 kHz
1 A 200 Hz ± 1.1 mA ± 1.3 mA ± 1.4 mA
5 kHz
100 mA
10 Hz
100 mA
3 Hz
± 110 µA ± 130 µA ± 140 µA1 kHz 200 Hz
5 kHz 200 Hz Q
100 µA 1 kHz
10 mA 200 Hz
± 4.1 µA
1 mA 1 kHz ± 4.7 µA ± 4.9 µA ± 5 µA
10 mA
1 kHz
± 11 µA ± 13 µA ± 14 µA
5 kHz
Keysight Truevolt Series Operating and Service Guide
519
Performance Verification