ACI Values for 34465A and 34470A DMMs
Input Error from Nominal
Current, rms Frequency Range Filter
Quick
Check
24 hours 90 days 1 year 2 years
1 mA
1 kHz
1 mA 200 Hz
Q
± 1.1 µA ± 1.3 µA ± 1.4 µA
5 kHz
100 µA
1 kHz
100 µA 200 Hz ± 0.110 µA ± 0.130 µA ± 0.140 µA
5 kHz
Q: Quick performance verification test point.
Performance Verification
520