Stratos Multi E401N
142
Error no. Type Message text Notes/solution
A023 ORP Diff HI Value above configured monitoring limit
A024 ORP Diff HI_HI Value above configured monitoring limit
A045 pH voltage diff range pH voltage difference. Max/min device limits exceeded:
- Check both pH voltage values.
- Check sensor/cable connections.
A046 pH Voltage Diff
LO_LO
Value below configured monitoring limit
A047 pH Voltage Diff LO Value below configured monitoring limit
A048 pH Voltage Diff HI Value above configured monitoring limit
A049 pH Voltage Diff HI_HI Value above configured monitoring limit
A200 Param. Calculation
block
Check the configuration:
Parameter Setting
System Control
Calculation Blocks
Conductivity (Contacting)
Error no. Type Message text Notes/solution
C001 Configuration Data
Loss
Data error in the configuration:
Fully reconfigure the device.
C008 Adjustment Data Error in the adjustment data:
Switch device off (approx. 10 s).
If the message persists, send in the device.
C009 Firmware Error Error in the firmware:
Switch device off (approx. 10 s).
Reload the firmware. ➜
Firmware Update (FW-E106), p. 213
If the message persists, send in the device.
C010 Conductivity Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect cell constant set.
Parameter Setting
[II] Analog Cond
Sensor Data
C011 Conductivity LO_LO Value below configured monitoring limit
C012 Conductivity LO Value below configured monitoring limit
C013 Conductivity HI Value above configured monitoring limit
C014 Conductivity HI_HI Value above configured monitoring limit
C015 Temperature Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect temperature probe selected.
Parameter Setting
[II] Analog Cond
Sensor Data
C016 Temperature LO_LO Value below configured monitoring limit
C017 Temperature LO Value below configured monitoring limit
C018 Temperature HI Value above configured monitoring limit
C019 Temperature HI_HI Value above configured monitoring limit
C020 Resistivity Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect cell constant set.
Parameter Setting
[II] Analog Cond
Sensor Data
C021 Resistivity LO_LO Value below configured monitoring limit