Stratos Multi E401N
143
Error no. Type Message text Notes/solution
C022 Resistivity LO Value below configured monitoring limit
C023 Resistivity HI Value above configured monitoring limit
C024 Resistivity HI_HI Failure: Value above configured monitoring limit
C025 Concentration Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect cell constant set.
Parameter Setting
[II] Analog Cond
Sensor Data
C026 Concentration LO_LO Value below configured monitoring limit
C027 Concentration LO Value below configured monitoring limit
C028 Concentration HI Value above configured monitoring limit
C029 Concentration HI_HI Value above configured monitoring limit
C040 Salinity Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, incorrect cell constant set.
Parameter Setting
[II] Analog Cond
Sensor Data
C041 Salinity LO_LO Value below configured monitoring limit
C042 Salinity LO Value below configured monitoring limit
C043 Salinity HI Value above configured monitoring limit
C044 Salinity HI_HI Value above configured monitoring limit
C045 Conductance Range Value above range limit. Possible causes: Sensor not/incorrectly connected,
incorrect sensor for range, cable faulty (short circuit).
C060 Sad Sensoface
Polarization
The sensor is polarized. Sensor is not suitable for the range or the process
medium. Connect a suitable sensor.
C062 Sad Sensoface
Cell Constant
Incorrect cell constant set, incorrect adjustment: Repeat calibration/
adjustment. Replace the sensor as required.
C070 TDS Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, incorrect cell constant set.
C071 TDS LO_LO Value below configured monitoring limit
C072 TDS LO Value below configured monitoring limit
C073 TDS HI Value above configured monitoring limit
C074 TDS HI_HI Value above configured monitoring limit
C090 par USP Limit The configured USP limit was exceeded.
C091 par Reduced USP Limit The configured reduced USP limit was exceeded.
C110 CIP Counter Configured number of CIP cycles exceeded: As required, calibrate/adjust
sensor or replace.
C111 SIP Counter Configured number of SIP cycles exceeded: As required, calibrate/adjust
sensor or replace.
C113 Sensor Operating
Time
Sensor operating time exceeded: Replace the sensor.
C120 Wrong Sensor If sensor check is active: An impermissible sensor is connected to the
device: Connect correct sensor or disable function.
C121 Sensor Error
(Factory Settings)
The digital sensor indicates an error. The sensor no longer works properly.
Replace the sensor.
C122 Sensor Memory (Cal
Data)
The digital sensor indicates an error. The calibration data are defective: Re-
calibrate/readjust sensor.