Stratos Multi E401N
144
Error no. Type Message text Notes/solution
C123 New Sensor,
Adjustment Required
A new digital sensor was connected. It must now be adjusted.
C124 Sensor Date The date of the sensor is implausible.e.g., calibration data “from the future”:
Check and, as necessary, adjust the configuration.
C204 Info Cal: Sensor Unstable The drift criterion was not adhered to during calibration. Possible causes:
Improper calibration. Sensor cable/connection faulty. Sensor worn out.
Check the sensor and calibration and repeat as required. Otherwise, replace
the sensor.
C205 Info Cal: Sensor Failure Replace the sensor.
Cond/Cond Calculation Block
Error no. Type Message text Notes/solution
E001 Configuration Data
Loss
Data error in the configuration:
Fully reconfigure the device.
E010 Conductivity Diff
Range
Conductivity value difference. Max/min device limits exceeded:
- Check both conductivity values.
- Check sensor/cable connections.
E011 Conductivity Diff
LO_LO
Value below configured monitoring limit
E012 Conductivity Diff LO Value below configured monitoring limit
E013 Conductivity Diff HI Value above configured monitoring limit
E014 Conductivity Diff
HI_HI
Value above configured monitoring limit
E015 Temperature Diff
Range
Temperature value difference. Max/min device limits exceeded:
- Check both temperature values.
- Check sensor/cable connections.
E016 Temperature Diff
LO_LO
Value below configured monitoring limit
E017 Temperature Diff LO Value below configured monitoring limit
E018 Temperature Diff HI Value above configured monitoring limit
E019 Temperature Diff
HI_HI
Value above configured monitoring limit
E020 Resistivity Diff Range Resistivity difference. Max/min device limits exceeded:
- Check both resistance values.
- Check sensor/cable connections.
E021 Resistivity Diff LO_LO Value below configured monitoring limit
E022 Resistivity Diff LO Value below configured monitoring limit
E023 Resistivity Diff HI Value above configured monitoring limit
E024 Resistivity Diff HI_HI Value above configured monitoring limit
E030 RATIO Range Ratio. Min/max device limits exceeded:
- Check both conductivity values.
E031 RATIO LO_LO Value below configured monitoring limit
E032 RATIO LO Value below configured monitoring limit
E033 RATIO HI Value above configured monitoring limit
E034 RATIO HI_HI Value above configured monitoring limit
E035 PASSAGE Range Passage. Min/max device limits exceeded:
- Check both conductivity values.