Stratos Multi E401N
145
Error no. Type Message text Notes/solution
E036 PASSAGE LO_LO Value below configured monitoring limit
E037 PASSAGE LO Value below configured monitoring limit
E038 PASSAGE HI Value above configured monitoring limit
E039 PASSAGE HI_HI Value above configured monitoring limit
E045 REJECTION Range Rejection. Min/max device limits exceeded:
- Check both conductivity values.
E046 REJECTION LO_LO Value below configured monitoring limit
E047 REJECTION LO Value below configured monitoring limit
E048 REJECTION HI Value above configured monitoring limit
E049 REJECTION HI_HI Value above configured monitoring limit
E050 DEVIATION Range Deviation. Min/max device limits exceeded:
- Check both conductivity values.
E051 DEVIATION LO_LO Value below configured monitoring limit
E052 DEVIATION LO Value below configured monitoring limit
E053 DEVIATION HI Value above configured monitoring limit
E054 DEVIATION HI_HI Value above configured monitoring limit
E055 Remaining Capacity
Range
The remaining capacity of the ion exchanger cannot be calculated.
E056 Degassed
Conductivity
Min/max device limits exceeded:
- Check both conductivity values.
E057 Ion Exchanger
Remaining Capacity
Remaining capacity of the ion exchanger < 20%:
Check ion exchanger; as required, replace the filter or the ion exchanger.
Remaining capacity of the ion exchanger 0%:
Replace the ion exchanger.
Replacement of the ion exchanger must be confirmed in the Maintenance
menu:
Maintenance
[CI] [CII] Cond-Cond Calculation
E060 pH Range pH range outside the permitted range of the VGB guideline:
- Check both conductivity values.
- Check choice of alkalizing agent.
- Check ion exchanger.
- Check both sensors/cables.
E061 pH LO_LO Value below configured monitoring limit
E062 pH LO Value below configured monitoring limit
E063 pH HI Value above configured monitoring limit
E064 pH HI_HI Value above configured monitoring limit
E200 Param. Calculation
Block
Check the configuration:
Parameter Setting
System Control
Calculation Blocks