Stratos Multi E401N
146
Conductivity (Inductive)
Error no. Type Message text Notes/solution
T001 Configuration Data
Loss
Data error in the configuration:
Fully reconfigure the device.
T008 Adjustment Data Error in the adjustment data:
Switch device off (approx. 10 s).
If the message persists, send in the device.
T009 Firmware Error Error in the firmware:
Switch device off (approx. 10 s).
Reload the firmware.
Firmware Update (FW-E106), p. 213
If the message persists, send in the device.
T010 Conductivity Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect cell factor set.
Parameter Setting
[II] Analog Condl
Sensor Data
T011 Conductivity LO_LO Value below configured monitoring limit
T012 Conductivity LO Value below configured monitoring limit
T013 Conductivity HI Value above configured monitoring limit
T014 Conductivity HI_HI Value above configured monitoring limit
T015 Temperature Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect temperature probe selected.
Parameter Setting
[II] Analog Condl
Sensor Data
T016 Temperature LO_LO Value below configured monitoring limit
T017 Temperature LO Value below configured monitoring limit
T018 Temperature HI Value above configured monitoring limit
T019 Temperature HI_HI Value above configured monitoring limit
T020 Resistivity Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect cell factor set.
Parameter Setting
[II] Analog Condl
Sensor Data
T021 Resistivity LO_LO Value below configured monitoring limit
T022 Resistivity LO Value below configured monitoring limit
T023 Resistivity HI Value above configured monitoring limit
T024 Resistivity HI_HI Value above configured monitoring limit
T025 Concentration Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect cell factor set.
Parameter Setting
[II] Analog Condl
Sensor Data
T026 Concentration LO_LO Value below configured monitoring limit
T027 Concentration LO Value below configured monitoring limit
T028 Concentration HI Value above configured monitoring limit
T029 Concentration HI_HI Value above configured monitoring limit