Error no. Type Message text Notes/solution
T040 Salinity Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect cell factor set.
Parameter Setting
[II] Analog Condl
Sensor Data
T041 Salinity LO_LO Value below configured monitoring limit
T042 Salinity LO Value below configured monitoring limit
T043 Salinity HI Value above configured monitoring limit
T044 Salinity HI_HI Value above configured monitoring limit
T045 Conductance Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, incorrect sensor for range, cable faulty (short circuit).
T060
Sad Sensoface
Primary Coil
Sensor faulty: Replace the sensor.
T061
Sad Sensoface
Secondary Coil
Sensor faulty: Replace the sensor.
T063
Sad Sensoface
Zero Point
Adjust the sensor zero point.
T064
Sad Sensoface
Cell Factor
Incorrect cell factor set, incorrect adjustment: Repeat calibration/adjust-
ment. Replace the sensor as required.
T070 TDS Range Measuring range exceeded. Possible causes: Sensor not/incorrectly
connected, cable incorrectly connected/faulty, range incorrectly
specified, incorrect cell factor set.
Parameter Setting
[II] Analog Condl
Sensor Data
T071 TDS LO_LO Value below configured monitoring limit
T072 TDS LO Value below configured monitoring limit
T073 TDS HI Value above configured monitoring limit
T074 TDS HI_HI Value above configured monitoring limit
T090 par USP Limit Configured USP limit was exceeded.
T091 par Reduced USP Limit The configured reduced USP limit was exceeded.
T110 CIP Counter Configured number of CIP cycles exceeded: As required, calibrate/adjust
sensor or replace.
T111 SIP Counter Configured number of SIP cycles exceeded: As required, calibrate/adjust
sensor or replace.
T113 Sensor Operating
Time
Sensor operating time exceeded: Replace the sensor.
T120 Wrong Sensor If sensor check is active: An impermissible sensor is connected to the
device: Connect correct sensor or disable function.
T121 Sensor Error
(Factory Settings)
The digital sensor indicates an error. The sensor no longer works properly.
Replace the sensor.
T122 Sensor Memory (Cal
Data)
The digital sensor indicates an error. The calibration data are defective: Re-
calibrate/readjust sensor.
T123 New Sensor,
Adjustment Required
A new digital sensor was connected. It must now be adjusted
T124 Sensor Date The date of the sensor is implausible.e.g., calibration data “from the future”.
Check and, as necessary, adjust the configuration.