Stratos Multi E401N
148
Error no. Type Message text Notes/solution
T205 Info Cal: Sensor Unstable The drift criterion was not adhered to during calibration. Possible causes:
improper calibration, sensor cable/connection faulty, sensor worn.
Check the sensor and calibration and repeat as required. Otherwise,
replace the sensor.
Oxygen
Error no. Type Message text Notes/solution
D001 Configuration Data
Loss
Data error in the configuration:
Fully reconfigure the device.
D008 Adjustment Data Error in the adjustment data:
Switch device off (approx. 10 s).
If the message persists, send in the device.
D009 Firmware Error Error in the firmware:
Switch device off (approx. 10 s).
Reload the firmware.
Firmware Update (FW-E106), p. 213
D010 Saturation %Air
Range
Max/min limits of range exceeded or sensor not/incorrectly connected, ca-
ble faulty.
D011 Saturation %Air
LO_LO
Value below configured monitoring limit
D012 Saturation %Air LO Value below configured monitoring limit
D013 Saturation %Air HI Value above configured monitoring limit
D014 Saturation %Air HI_HI Value above configured monitoring limit
D015 Temperature Range Max/min limits of range exceeded or sensor not/incorrectly connected, ca-
ble faulty.
D016 Temperature LO_LO Value below configured monitoring limit
D017 Temperature LO Value below configured monitoring limit
D018 Temperature HI Value above configured monitoring limit
D019 Temperature HI_HI Value above configured monitoring limit
D020 Concentration Range Max/min limits of range exceeded, sensor not/incorrectly connected,
cable faulty.
D021 Concentration LO_LO Value below configured monitoring limit
D022 Concentration LO Value below configured monitoring limit
D023 Concentration HI Value above configured monitoring limit
D024 Concentration HI_HI Value above configured monitoring limit
D025 Partial Pressure Range Max/min limits of range exceeded, sensor not/incorrectly connected,
cable faulty.
D045 Saturation %O2
Range
Max/min limits of range exceeded, sensor not/incorrectly connected,
cable faulty.
D046 Saturation %O2
LO_LO
Value below configured monitoring limit
D047 Saturation %O2 LO Value below configured monitoring limit
D048 Saturation %O2 HI Value above configured monitoring limit
D049 Saturation %O2 HI_HI Value above configured monitoring limit
D060
Sad Sensoface
Slope/
- Readjust sensor.
- Check/refill electrolyte.
- SE740: Replace membrane cap.
- Replace sensor.