Error no. Type Message text Notes/solution
D061 Neutral Sensoface
Zero Point
- Readjust sensor.
- Check/refill electrolyte.
- SE740: Replace membrane cap.
- Replace sensor.
D062 par
Sad Sensoface
Sensocheck
- Readjust configured sensor.
- Replace sensor.
D063
Sad Sensoface
Response Time
- Check/refill electrolyte.
- SE740: Replace membrane cap.
- Replace sensor.
D064
Sad Sensoface
Cal Timer
Calibration timer has expired:
- Check the calibration timer settings.
- Carry out calibration/adjustment.
D070 par
Sad Sensoface
Wear
Sensor is worn (100%):
- Readjust sensor.
- Check/refill electrolyte.
- Replace sensor.
D071 par
Sad Sensoface
Membrane
For SE740 optical oxygen sensor only: Replace membrane cap.
D080 Sensor Current Range - Check polarization voltage:
Parameter Setting
[I] [II] … Oxy
Sensor Data
- Refill electrolyte.
- Recalibrate/readjust.
D081 O2 Measurement OFF
(Temp)
For SE740 optical oxygen sensor only: The oxygen measurement was auto-
matically stopped because the temperatures are too high.
CIP/SIP Counters, p. 94
D110 CIP Counter Configured number of CIP cycles exceeded: As required, recalibrate/read-
just sensor or replace.
D111 SIP Counter Configured number of SIP cycles exceeded: As required, recalibrate/read-
just sensor or replace.
D112 Autoclaving Counter Configured number of autoclaving cycles exceeded: As required, recali-
brate/readjust sensor or replace.
D113 Sensor Operating
Time
Sensor operating time exceeded.
Replace the sensor.
D114 Membrane Body Re-
placement
Replacement of the membrane body required.
Replacement of the membrane body must be confirmed in the Mainte-
nance menu.
Maintenance Functions for Oxygen Sensors, p. 134
D120 Wrong Sensor If sensor check is active: An impermissible sensor is connected to the
device.
- Replace sensor.
- Change the process variable.
D121 Sensor Error
(Factory Settings)
The digital sensor indicates an error. The sensor no longer works properly.
Replace the sensor.
D122 Sensor Memory (Cal
Data)
The digital sensor indicates an error. The calibration data are defective:
Recalibrate/readjust sensor.
D123 New Sensor,
Adjustment Required
A new digital sensor was connected. It must now be adjusted.
D124 Sensor Date The date of the sensor is implausible.
e.g., calibration data “from the future”: Check and, as necessary, adjust the
configuration.
D200 Temp O2 Conc/Sat The temperature is outside the valid range for oxygen concentration/
saturation.