Stratos Multi E401N
150
Error no. Type Message text Notes/solution
D201 Cal: Temperature The calibration temperature is impermissible: Check the calibration temper-
ature. Note the information in the Calibration chapter.
Process Variable Oxygen Calibration/Adjustment, p. 121
D205 Info Cal: Sensor Unstable The drift criterion was not adhered to during calibration. Possible causes:
improper calibration, sensor cable/connection faulty, sensor worn. Check
the sensor and calibration and repeat as required. Otherwise, replace the
sensor.
Oxy/Oxy Calculation Block
Error no. Type Message text Notes/solution
H001 Configuration Data
Loss
Data error in the configuration:
Fully reconfigure the device.
H010 Saturation %Air Diff
Range
Saturation value difference. Max/min device limits exceeded:
- Check both saturation values.
- Check sensor/cable connections.
H011 Saturation %Air Diff
LO_LO
Value below configured monitoring limit
H012 Saturation %Air Diff
LO
Value below configured monitoring limit
H013 Saturation %Air Diff
HI
Value above configured monitoring limit
H014 Saturation %Air Diff
HI_HI
Value above configured monitoring limit
H015 Temperature Diff
Range
Temperature value difference. Max/min device limits exceeded:
- Check both temperature values.
- Check sensor/cable connections.
H016 Temperature Diff
LO_LO
Value below configured monitoring limit
H017 Temperature Diff LO Value below configured monitoring limit
H018 Temperature Diff HI Value above configured monitoring limit
H019 Temperature Diff
HI_HI
Value above configured monitoring limit
H020 Conc. (Liquid) Diff
Range
Concentration value difference. Max/min device limits exceeded:
- Check both concentration values.
- Check sensor/cable connections.
H021 Conc. (Liquid) Diff
LO_LO
Value below configured monitoring limit
H022 Conc. (Liquid) Diff LO Value below configured monitoring limit
H023 Conc. (Liquid) Diff HI Value above configured monitoring limit
H024 Conc. (Liquid) Diff
HI_HI
Value above configured monitoring limit
H045 Saturation %O2 Diff
Range
Saturation value difference. Max/min device limits exceeded:
- Check both saturation values.
- Check sensor/cable connections.
H046 Saturation %O2 Diff
LO_LO
Value below configured monitoring limit
H047 Saturation %O2 Diff
LO
Value below configured monitoring limit
H048 Saturation %O2 Diff
HI
Value above configured monitoring limit