EasyManua.ls Logo

Knick Stratos Multi E401N - Page 60

Knick Stratos Multi E401N
222 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Stratos Multi E401N
60
Usage: Rinse Contact
Please note when configuring the “Rinse Contact” function
The “function check (HOLD) mode (e.g. during parameter setting) delays the execution of the
rinse contact function.
Up to 3 rinse functions (contacts K1 ... K3) can be configured independently of each other.
The individual rinse functions are not synchronized with each other.
Configuring the Rinse Contact
01. Open Inputs and Outputs
Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: Rinse Contact
04. Select contact type (e.g. “N/O”).
05. Specify rinsing interval.
06. Specify rinse duration.
07. Specify rinse/meas. lead time.
Note: Function check (HOLD) is active during the defined “Lead Time...
08. Logbook entry: off/on
Usage: USP Output
Can be activated when using a conductivity sensor and the USP function USP Function, p. 81
01. Open Inputs and Outputs
Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: USP Output
04. Assign USP channel.
05. Set the contact parameters.
Usage: Sensoface
Sensoface messages can be output via a relay contact.
If using two sensors, the Sensoface messages can be assigned to different contacts:
01. Open Inputs and Outputs
Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: Sensoface Channel
04. Select channel.
Contact K1 (Admin.)
Channel
Back
Usage
Contact Type
ON Delay
OFF Delay
Sensoface Channel
Memosens pH
N/O..
I
Memosens pH
Memosens Cond
II
0 s
0 s
05. Set the contact parameters.
06. Using the left softkey: Back, go to the Relay Contacts menu and configure the second contact.
See also
Sensocheck and Sensoface, p. 151

Table of Contents