Stratos Multi E401N
90
Selecting a Second Memosens Oxygen Sensor
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [II]
Selection of a second Memosens oxygen sensor connected to the MK-MS 095N measuring module:
Module: MK-MS
Parameter: Oxygen
Mode: Memosens
Functionality: Amperometric
Adjustable parameters for Memosens oxygen sensors
Parameter Setting
[I] [II] Memosens Oxy
Input Filter Pulse Suppression Enable/disable suppression of interference pulses.
Sensor Data
➜
Sensor Data, p. 92
Measure in Liquids, gases
Sensoface Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Option to enter individual limits for monitoring separate pa-
rameters.
Disable Sensocheck sensor monitoring or select whether
Sensocheck should generate failure or maintenance required
messages.
Option to enter individual limits before a message for response
time, sensor wear, sensor operating time, SIP counter is trig-
gered.
Cal Presettings Presetting of the calibration mode and calibration timer.
➜
Calibration Presettings, p. 95
Pressure Correction Manual input of pressure during measurement and calibration.
With TAN option FW-E051: automatic pressure correction by external pressure transmit-
ter. ➜
Pressure Correction, p. 95
Salinity Correction
Salinity, chlorinity, conductivity ➜ Salinity Correction, p. 96
Messages Enable/disable messages for the separate process variables, or specify individual limits.
➜
Messages, p. 96
Selecting an SE740 Optical Oxygen Sensor (LDO)
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [I]
Selection of the SE740 optical oxygen sensor connected to the RS-485 interface (terminals 1 … 6):
Parameter: Auto or oxygen
Mode: Other digital
Functionality: SE740
Note: The voltage at terminal 6 (Power Out) is automatically set to 14V when using the SE740 optical
oxygen sensor. The Power Out menu item is not available.
Adjustable parameters for the SE740 optical oxygen sensor
Parameter Setting
[I] Digital Oxy
Input Filter Pulse Suppression Enable/disable suppression of interference pulses.
Sensor Data
➜
Sensor Data, p. 92
Measure in Liquids, gases
Sensoface Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Select whether a failure or maintenance required message is
triggered if the range of the Stern-Volmer constant and phase
angle is exceeded.
Disable Sensocheck sensor monitoring or select whether
Sensocheck should generate failure or maintenance required
messages.
Option to enter individual limits before a message for sensor
wear, sensor operating time, CIP counter, O
2
measurement with
CIP/SIP, autoclaving counter is triggered.