Stratos Multi E401N
91
Adjustable parameters for the SE740 optical oxygen sensor
Parameter Setting
[I] Digital Oxy
Cal Presettings Presetting of the calibration mode and calibration timer.
➜
Calibration Presettings, p. 95
Pressure Correction Manual input of pressure during measurement and calibration.
With TAN option FW-E051: automatic pressure correction by external pressure transmit-
ter. ➜
Pressure Correction, p. 95
Salinity Correction
Salinity, chlorinity, conductivity ➜ Salinity Correction, p. 96
Messages Enable/disable messages for the separate process variables, or specify individual limits.
➜
Messages, p. 96
Selecting an Analog Oxygen Sensor
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [II]
Selection of an oxygen sensor connected to the MK-OXY046N measuring module:
Module: MK-OXY
Mode: Analog
Adjustable parameters for analog oxygen sensor
Parameter Setting
[II] Analog Oxy
Input Filter Pulse Suppression Enable/disable suppression of interference pulses.
Sensor Data
➜
Sensor Data, p. 92
Measure in Liquids, gases
Sensor Type Standard or other
Temperature Probe NTC 22kΩ, NTC 30kΩ
Sensor Polarization Automatic or individual
If you select “Individual”, separate values for polarization
can be entered during measurement and calibration.
Membrane Compensation If “Other Sensor Type” is selected
Sensoface Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Option to enter individual limits for monitoring zero point
and slope.
Disable Sensocheck sensor monitoring or select whether
Sensocheck should generate failure or maintenance re-
quired messages.
Option to enter individual limits before a message for re-
sponse time is triggered.
Cal Presettings Presetting of the calibration mode and calibration timer.
➜
Calibration Presettings, p. 95
Pressure Correction Manual input of pressure during measurement and calibration.
With TAN option FW-E051: automatic pressure correction by external pressure transmit-
ter. ➜
Pressure Correction, p. 95
Salinity Correction
Salinity, chlorinity, conductivity ➜ Salinity Correction, p. 96
Messages Enable/disable messages for the separate process variables, or specify individual limits.
➜
Messages, p. 96