6 OPERATION
100 TITAN VERSA Operations and Maintenance Manual LACO Technologies
Counters
Cycle Count Displays the current cycle count.
Valve Maint. Counter At over 500,000 cycles the valves should be replaced.
Turbo Pump Hours Displays current turbo pump hours. Reset only with a new or refurbished turbo pump.
Turbo Maint. Hours Turbo pumps should be serviced every 17,200 hours.
Filament 1 / 2 Hours Displays the current hours for filaments 1 and 2. When filaments are replaced the hour
counters should be reset to 0.
Detector hours Display the current total detector hours.