8 MAINTENANCE AND SERVICE
LACO Technologies TITAN VERSA Operations and Maintenance Manual 137
Pump Fan Filter Tower Configuration
TITAN VERSA Leak Components
Part Number
Description
CM112498/L-7/4
Calibrated Leak
PFP0488E1
Temperature Board
LSLC0001
Calibrated Leak Standard Calibration, Single Point
LSLC0003
Sniffer Calibrated Leak Standard Calibration, Single Point
Recalibration
NOTICE: Leak Standard Calibration Frequency
LACO recommends that the internal calibrated leak standard be calibrated at
least every two years.
Sniff calibrated leaks and higher leak rate external leaks (> 1.0 e-6 mbar*L/sec)
should be calibrated yearly, or more frequently if the depletion rates are large.
Depletion rate is indicated on the calibrated leak identification label.
Return leak standards to LACO Technologies calibration lab for recalibration.