6 OPERATION
82 TITAN VERSA Operations and Maintenance Manual LACO Technologies
Calibration Method Select option of either Internal (Vacuum, Mass 4 only) or External calibration per Table 33.
Leak Rate Enter the leak standard leak rate.
Units Select the leak standard units.
Date Enter the month and year of the leak standard calibration in the MMYYYY format.
Cal Temp Enter the leak standard calibration temperature in Celsius.
Temp Coefficient Enter the temperature coefficient in 0.1 % precision.
Depletion Rate Enter the depletion rate. Round to nearest whole percentage number.
Current Temperature Enter the estimated current leak standard temperature in degrees Celsius (only for
External leaks).
Calibration Start
Cal Start Method There are three options for calibration start.
• Cal at Start (not recommended but available)
• Cal Check at Start (This runs both a cal check and an autocalibration)