EasyManua.ls Logo

LACO Technologies TITAN VERSA C - General Use Guidelines; Calibration

LACO Technologies TITAN VERSA C
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 OPERATION
LACO Technologies TITAN VERSA Operations and Maintenance Manual 45
Calibration
NOTICE: Calibration Frequency
After the leak detector has been powered on for 20 minutes, the calibration
warning screens appears. It is recommended operators calibrate the unit 20
minutes after each time the unit is powered on.
NOTICE: Calibration Settings
Verify all calibrated leak settings are correct before performing a calibration.
Refer to Calibration Settings.
NOTICE: Leak Standard Calibration Frequency
LACO recommends the internal calibrated leak standard be calibrated at least
every two years.
Sniff calibrated leaks and higher leak rate external leaks (> 1.0 e-6 mbar*L/sec)
should be calibrated yearly, or more frequently if the depletion rates are large.
The TITAN VERSA Leak Detector offers both internal and external calibrations as
outlined in the table below.
Calibration Types per Mode and Mass
ID
Test Mode
Test Gas Mass
Calibration Type
1
Vacuum
Helium 4
Internal
2
Vacuum
External
3
Sniff
4
Vacuum
Helium 3
5
Sniff
6
Vacuum
Hydrogen
7
Sniff

Table of Contents

Related product manuals