6 OPERATION
LACO Technologies TITAN VERSA Operations and Maintenance Manual 81
Calibrated Leak Standard
Table 33 below outlines the seven sets of parameters for calibrated leak settings per
Mass and Test Mode selected by the user. See Calibration for information on
performing the different types of calibrations.
Calibration Method per Mode and Mass
Internal calibration cycles are fully automated using the internal leak standard. For
external calibrations, the user must follow the on-screen prompts to introduce and
remove the leak standard at certain time. See Section 6.4.1 for more information on
performing the different types of calibrations. Reference the leak standard label to
enter the parameters below.