6 OPERATION
76 TITAN VERSA Operations and Maintenance Manual LACO Technologies
To access Data Log settings, go to either Test Settings or Test Quick-Access.
The three data logging methods are:
Test Summary
Stream
Cycle Combo (Test Summary and Stream during test cycle)
Test Summary Method The Test Summary Method logs a row of test summary data at the end of a test cycle or
aborted test cycle.
The first time a Test summary log occurs, a new TestSummary.csv file is created
and appears on the USB drive. Subsequent test cycles add additional rows to the
existing csv file.
Enter any Test ID data that is desired before running the leak test (see Test ID ).
Entered data will be logged.
The Test Summary Method is ideal for users requiring traceability for their leak
test application.
To enable the Test Summary Method:
4. Select Test Summary from the log method and press the Data Log Enable button.
5. When in a test cycle is in process, the Data Log icon will flash with a T symbol.
Additionally, the bottom the test screen will read “Data Logging Active”.
6. The data is logged at the end of the test cycle.
Test Summary Field Description
Data Field
Comments
Date
Time
Leak Rate
Final Leak rate
Pressure
Final Pressure reading
Method
Vacuum or Sniff
Vac Mode
Ultra, Fine, or Gross
Result
Pass, Fail or Incomplete
Figure 10: Example of Test Summary csv File from Excel
Operator
Part Number
Serial Number
LOT #
TB
Date
Time
Leak Rate
Pressure
Method
Mode
Pass
Operator
Part Num
Serial #
Lot Num
Operator
1/1/2019
15:45:40
2.17E-06
0.174
Vacuum
Ultra
Fail
Operator
Part Num
Serial #
Lot Num
Operator
1/1/2019
15:55:59
2.78E-11
0.001
Vacuum
Ultra
Pass