6 OPERATION
102 TITAN VERSA Operations and Maintenance Manual LACO Technologies
Background Cleanup When Background Cleanup is selected, the main screen appears and a background
cleanup process runs. This process consists of running the detector for one hour while
the unit undergoes an automatic, continuous cycle and vent process. This process can
be viewed either from the Test screen or the Manual Control screen, and is ideal for
applications where the background requires improvement or when new valves are
installed on the valve block.