8 MAINTENANCE AND SERVICE
138 TITAN VERSA Operations and Maintenance Manual LACO Technologies
Replacement
1. Remove leak detector cover (see Removing Covers).
2. Loosen retaining screws with a 2.5 mm hex wrench and rotate leak unit until it
is free.
3. Disconnect the temperature sensor from the calibrated leak using a 2.5 mm Hex
wrench.
4. Connect Temp Sensor to replacement leak and insert into retaining device.
5. Update the settings of the internal calibrated leak.