6 OPERATION
LACO Technologies TITAN VERSA Operations and Maintenance Manual 83
Manual (after 20 minutes of warm-up)
Cal Check Start Method Refer to Calibration Check (Cal Check) for more information on this function. The default
manual method allows the user to run a cal check whenever desired. Automatic method
will automatically run a cal check when either the Hours or cycles have been reached.
Cal Check Auto Hours Enter desired hours before performing cal check.
Cal Check Auto Cycle Enter desired cycles before performing cal check.
NOTICE: Use caution when using the Machine factors as they can give
incorrect readings if used improperly. The factors will adjust all leak test
readings. Only use machine factors when external calibrations have been used
to verify correct readings.
Machine Factor-Vacuum The machine factor is only used in split flow applications.
Machine Factor-Sniff Enter desired machine factor for sniff mode.