5 INSTALLATION
LACO Technologies TITAN VERSA Operations and Maintenance Manual 23
Figure 2: Bronze Filter Installation
The filter should be cleaned with clean, dry air every three months and replaced every
18 months.
5.2.1.2. Optimized Vacuum Test Performance Recommendations
Verify test setup has proper hoses, connections, and valves in accordance with
good vacuum practice.
Use pipe with a diameter equal to the diameter of the detector's inlet. The pipes
should be as short as possible and completely sealed.
Do not use plastic hoses such as compressed air pipes.
Check that the connected part/installation is impermeable to tracer gas.
Test only clean, dry parts or installations with no trace of water, vapor, paint,
detergent, or rinsing products.
Test that the entire line is completely sealed when the detector is attached to the
pumping circuit to ensure the connections (pump, pipe, valves, etc.) are correct.