6 OPERATION
LACO Technologies TITAN VERSA Operations and Maintenance Manual 43
3. Verify proper operation voltage per the product label and switch on the device with
the power entry module. The TITAN VERSA will energize and go into startup phase.
4. During startup, the turbo pump will spin up to full speed (1500 Hz). The startup
time for the turbo pump ranges from two to five minutes. After the turbo pump
reaches full speed, the Mass Spectrometer (Mass Spec) will verify filament function
and proper emission.
If the leak detector has not been used for 15 days, the unit will undergo a storage
delay setup.
5. After startup is complete, the test screen appears, and the leak detector status will
read “Standby”.
NOTICE: Allow the device to warm up for 20 minutes before making exact
measurements or executing a calibration.