6 OPERATION
50 TITAN VERSA Operations and Maintenance Manual LACO Technologies
5. Press the Calibrate icon.
6. Press Start.
7. Verify External Calibrated leak setting configurations. Press Continue to start the
calibration or Edit to change the calibrated leak valve settings.
8. Confirm sniffer probe is connected and leak rate is stable. Keep sniffer probe
inserted.
9. Calibration proceeds through peak find and peak measure.
10. Dialog prompts user to remove sniffer probe. After probe is removed, press
Continue.