LUMENIS VersaPulse PowerSuite - Page 141

LUMENIS VersaPulse PowerSuite
178 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9HUVD3XOVHPowerSuite6HULHV6HUYLFH0DQXDO 7528%/(6+227,1*
5(9% 
TP6 LampVolts Frequency to Command HVPS B+ Voltage on Cap Bank
TP7 CROWB Crowbar Drive Signal to HVPS (Normally LO)
TP8 CBOK Readback of Crowbar Signal from HVPS (Normally HI)
TP9 HVPS Analog Readback Frequency of Selected Signal(s) Returned from HVPS
(See TP16)
TP11 RemoteRX Receive Serial Data from Remote Control Panel
TP12 D+5V +5V Drive Signal to Laser Diode Module (ON/OFF
Control)
TP13 DWIP Digital Pot Voltage for Laser Diode (Intensity Control)
TP14 RemoteTX Transmit Serial Data to Remote Control Panel
TP15 SFDIF Safety Energy Detector Signal after Zero Correction Stage
TP16 HVPS Analog Readback Signal(s) Returned from HVPS Converted to DC Level
TP17 SFZER Safety Energy Detector Signal after Differential Amplifier
Stage
TP18 GND Ground Reference Point
TP19 GND Ground Reference Point
TP20 GND Ground Reference Point
TP21 MNZER Main Energy Detector after zero correction
TP22 GTME Greater than Min. Energy (Strobes HI with each laser
pulse)
TP23 MENERG Main Energy Signal after Peak/Hold Detector (Scope)