Installation and User Manual
89
7.3.5
Programming parameters summary table
Table 1 Programming menu parameters (for use by INSTALLATION PERSONNEL)
Display
Function Range Config.
U.M.
Default
Availability
Man Auto
IP DELAY
Delay Time
0,0 ÷ 300,0 s 0,0
O O
DLY RTEN
Disables retraction during
delay time
IP: yes / no
IP+POS: all / IP / POS / no
Yes
O X
RET t1
Self-timed reload at the C0
0.0 …5.0 0.0 s
O O
RET t2
Self-timed reload at the C0
trigger: reload period after
0.0 …5.0 0.0 s
O O
CONTROLS
Number of controls
C0 / C1 / C2 / C3 / C4 C4
O X
C-1 EN
Enable control C-1
yes / no No
O X
C-1
Control C-1 value
0 to – Full scale µm / “
- Full
O X
TIP t1
Self-timed TIP: wait time 0.1…30.0 30.0 s
O O
TIP t2
Self-timed TIP: C1 disable
0.1…5.0 0.1 s
O O
FBK STEP
Pulse feedback correction
step
0.00001/
0.00002/
0.00004/
“ 0.00004
O X
FBK RST
Automatic reset of the
Pulse Feedback value at
cycle end (InProcess
yes / no No
O X
Zero Adjust correction mode
ADJ INCR
Use this function to vary the
direction of rotation of the
knob used to modify the zero
adjust value
CCW: Increase by rotating
counter-clockwise
CW: Increase by rotating
clockwise
CCW
O O
ADJ SPED
Knob acceleration used to
modify the Zero Adjust
fast / slow3 / slow2 /
slow1
fast
O O
MEM DAMP
Damping factor
No damp
damp 1
damp 2
damp 3
damp 3
O O
DAMP TH.
Damping activation
threshold.
20...Full scale (scale
200-0-40) or
10… Full scale or
5...Full scale (scale 10-
0-2)
Full
scale
or 80
(scale
200-0-
µm
O O
0.0002...Full scale
full
inch
O O