EasyManua.ls Logo

MICRO-EPSILON ILD1700-2 - Menu, Setting the Parameters

MICRO-EPSILON ILD1700-2
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 37
Operation
optoNCDT 1700
6.5 Menu, Setting the Parameters
The sensor parameters can be set in setup mode using the function/enter and select/zero keys.
Measurement value
output
Measurement
mode
function
enter
Save
parameters
Current output
4 ... 20 mA
Voltage outp.
0 ... 10 VDC
RS422
Measuring frequency
function
enter
Averaging number N
function
enter
flashes green
LED state
1
2.5 kHz
1/2
1.25 kHz
1/4
625 Hz
1/8
312.5 Hz
1
(Median 3)
4
(Median 5)
128
(Median 9)
32
(Median 7)
hold last
value
3 mA
10.2 VDC
function
enter
LED state
redflashes
B CA
Error analog
red (flashes)
LED output
red green
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
select
zero
LED speed yello
red (flashes)
red green
LED avg yello
red (flashes)
red green
function
enter
function
enter
function
enter
function
enter
red (flashes)
LED output
red
IMPORTANT!
Parameters for
- Output type
(„Measurement value
output“)
- Measurement
frequency
- Average number
- Analog error
The system returns to
measurement mode
without saving the
parameters:
- if 15 seconds elapse
after the last press of
the function/en-
ter
key.
- if 30 seconds elapse
after the last press of
the select/zero
key.

Table of Contents