Appendix B TriStar II 3020
B-8 302-42828-01 - Dec 2013
2493- An entry is required for this field.
2494- Value is out of the valid range.
2495- Value is out of the valid range. Enter a value between [value] and [value].
2496- Invalid number.
2497- This field contains an invalid character.
2498- The requested change to the Sample’s status is invalid at this time.
2499- Sequence number must contain at least 3 digits.
2586- Sample file [name] has no pressure table entries.
Cause: This field requires a valid entry to proceed.
Action: Enter or select an appropriate value.
Cause: The entered value in the highlighted field is outside the valid range of
values.
Action: Check the entry and enter or select an appropriate value.
Cause: An invalid number entered in the highlighted field.
Action: Check the entry and enter or select a valid number.
Cause: An invalid character was entered in the highlighted field.
Action: Check the entry and enter valid characters.
Cause: A request to change the file’s status, for example, from automatically
collected to manually entered could not be done.
Action: Contact a Micromeritics service representative if this error message
continues. Record the name of the sample file in which the problem
occurred.
Cause: An attempt was made to enter a sequence number that did not contain
at least three digits.
Action: Enter a sequence number that contains at least three digits.
Cause: An attempt was made to start an analysis with an empty target
pressure table.
Action: Add target pressures to the pressure table in the analysis conditions.