TriStar II 3020 Appendix B
302-42828-01 - Dec 2013 B-7
2486- Could not construct [name] report type. Program will terminate.
.2487- Could not start report generator. Error code [number]. Program will
terminate.
2488- File [file name] cannot be opened for editing. It is already in use.
2489- File [file name] cannot be opened for writing. It is already in use.
2490- No ‘.INI’ file present. Application will terminate.
2491- Highlighted fields contain errors. Please correct the errors before closing dia-
log box.
2492- This field’s entry is invalid.
Cause A: Full rights to the application’s folders and files is required.
Action A: Contact a system administrator to have full rights granted.
Cause B: An internal processing and/or hardware error has occurred.
Action B: Contact a Micromeritics service representative if this error message
continues.
Cause: The specified file is being used in another edit operation.
Action: Check the Windows list to locate the other edit session.
Cause: The specified file in a Save As operation is already open for edit.
Action: Select a different file for the Save As operation.
Cause: The ASCII file containing initialization information and system
options information used during program startup does not exist.
Action: Run the analyzer setup program, select Change analyzer setup and
create the control file used by the analyzer.
Cause: The highlighted fields contain invalid entries. The dialog box cannot
be closed until all errors are corrected.
Action: Check the entries, correct the errors and close the window.
Cause: The highlighted field contains an invalid entry.
Action: Check the entry and correct the error.