Appendix B TriStar II 3020
B-6 302-42828-01 - Dec 2013
2480- File [name] cannot be analyzed. It is currently being edited.
2481- Error accessing the sample information file [file name].
2482- File cannot be opened for writing. It is already in use.
2483- An analysis cannot be performed on [file name]. It is open for editing and con-
tains errors.
2484- The edit session for [file name] must be saved before the analysis. Save
changes and continue with the analysis? [Yes, No]
2485- The service test file has an invalid status and cannot be used for this analysis.
Cause: An attempt was made to start an analysis using a file that is open for
editing.
Action: Finish editing the file, save and close it then start the analysis.
Cause: An unexplained error prevented access to this file.
Action: The hard disk drive may be corrupt. Run diagnositics.
Cause: An attempt was made to open a file currently being used.
Action: Locate the application using the file (in the Micromeritics application,
use the Windows menu item to get a list of all windows, one of which
may contain this file).
Cause: An attempt was made to use a sample file containing errors that is
currently open.
Action: Go to the window containing the file, correct the errors and save it.
Cause: An attempt was made to start an analysis using a file that contains
unsaved changes and is open for editing.
Action: Select Yes to save the changes and proceed with the analysis. Select
No to cancel the analysis and continue editing the Sample
Information file.
Cause: The selected service test file has a status other than No Analysis.
Action: Select a different service test file or create a new one and use Replace
All to copy parameters from the file originally selected.