Appendix B TriStar II 3020
B-24 302-42828-01 - Dec 2013
4058- At least two data points are needed for Temkin calculations.
4059- Fewer than 2 points available for MP-Method calculations.
4060- Sample [file name] contains no data points.
4061- The t-curve must contain at least 2 points.
4062- Error during report preparation.
Cause A: Less than two data points have been selected for the Temkin report; at
least two are required.
Action A: Select Temkin points on the Collected Data window. If calculation
assignments are not being used, edit the Temkin Report options,
Absolute pressure range in the sample file.
Cause B: At least one point with a negative pressure value was selected to
include in the Temkin report.
Action B: Negative pressure points void Temkin calculations. Open the sample
file and click the Collected Data tab. Locate the negative values in
the Temkin column and deselect it.
Cause: At least two points are required for the MP-Method calculations.
Action: Edit the calculation assignments for the MP-Method report.
Cause: An attempt was made to save a sample without collected data as a t-
curve or alpha-S curve.
Action: Repeat the Save As t-curve or Save As alpha-S operation after
opening a sample that has collected data.
Cause: At least two points are required in a thickness curve definition.
Action: Edit the thickness curve.
Cause: An internal processing error has occurred.
Action: Contact a Micromeritics service representative if this error message
continues.