TriStar II 3020 Appendix B
302-42828-01 - Dec 2013 B-25
4063- The data requested on this report are not available. No subreports selected.
4067- No data points are within the range of pressures in the reference isotherm.
4068- No points were selected for the f-Ratio report.
4070- Unable to load deconvolution model [name].
4071- The selected pressures points do not form a valid set for deconvolution.
Cause: There is no information in the sample log to report.
Action: A sample file was selected which no instrument operations has been
used. Select a sample file with a status of Prepared, Preparing,
Analyzing or Complete to obtain a valid sample log report.
Cause: There are no collected data points within the range of pressures in the
reference isotherm.
Action: Select data points in the range of the reference isotherm or select a
more appropriate reference isotherm.
Cause: The collected data column for the f-Ratio report does not have any
points selected.
Action: Edit the collected data window and select points for the f-Ratio
report.
Cause: For some reason, the list of available models was corrupted,
therefore, the model selected could not be loaded for the
deconvolution.
Action: Exit the program and reinstall the software, then try again.
Cause: The data points selected for analysis do not contain enough
information to allow a DFT data reduction.
Action A: Edit data points in the table of the Collected Data window or select
another sample file. At least two points with strictly increasing
pressures and volumes adsorbed are required for a DFT Plus data
reduction.
Action B: Edit the pressure range on the DFT report options window.