4 SMART VACPREP CALIBRATION
Smart VacPrep > Unit [n] > Calibration
Various calibration procedures and service tests are included in the operating program. These procedures
and tests are designed to provide a service representative with instrument readouts, as well as to assist in
troubleshooting potential problems.
SMART VACPREP PRESSURE OFFSET
Smart VacPrep >Unit [n] > Calibration >Pressure Offset
Use to perform system calibration. This option evacuates the system and zeros the pressure transducer.
Disabled calibration options can be accessed only with the assistance of an authorized Micromeritics
service representative.
Click Start to begin the process.
SMART VACPREP SERVO VALVE CALIBRATION
Smart VacPrep >Unit [n] > Calibration >Servo Valve
4 Smart VacPrep Calibration
Smart VacPrep Operator Manual 067-42800-01
Dec 2013
4 - 1