4 Smart VacPrep Calibration
Use to calibrate the servo valve to the manifold pressure transducer. The servo valve should always be
recalibrated after a pressure calibration has been performed. The pressure transducer should be
calibrated before starting this calibration procedure.
1. Go to Smart VacPrep > Unit [n] > Calibration > Servo Valve.
2.
Click Start. The window closes when the calibration is complete. Click Cancel to stop the cal-
ibration process.
Field or Button Description
Cancel Discards any changes or cancels the current process.
Start Starts the operation.
Servo Valve Fields and Buttons Table
4 - 2 Smart VacPrep Operator Manual — 067-42800-01
Dec 2013