6-11 
Failure 
Description 
Possible Cause  Recommended Action 
increase the current of the inspiratory valve 
(Insp Valve Current
measured value of the calibration device under 
each valve current. If the collected AD value 
of the expiratory flow sensor at the time point 
with the measured value of the calibration 
device close to (smaller than) 180 L/min is 
greater than 4050, the measurement range of 
the expiratory flow sensor is abnormal. 
Replace the expiratory flow sensor. 
3.  Replace the VCM. 
Display 34 
The drive gas pressure is 
low. 
1.  Replace or connect a gas supply to keep the 
gas supply pressure within 280–600 kPa  (40 
psi–87 psi). 
2. 
If the gas supply is normal, check the gas 
supply pressure switch. 
Display 39  T
data into EEPROM. 
1.  Perform the calibration again. 
2.  Replace the VCM. 
Display 3B  The ACGO switch is in the 
ON position. 
1. 
Check whether the ACGO switch is set to 
OFF. 
2. 
Check the monitoring status of the ACGO 
switch on the screen of the anesthesia 
machine.