7-55 
Cause Confirmation 
Solution 
(3) If the problem persists, 
replace the monitoring and 
controlling assembly (refer 
 
7.4.11 Auxiliary O2/Air and HFNC Module Failures 
Cause Confirmation 
Solution 
Auxiliary O2/air 
or HFNC module 
startup failure 
(backlight of the 
segment display 
and flowmeter is 
off) 
1. Turn on the auxiliary 
O2/air or HFNC module, 
power on the main unit, and 
check for the Aux O2/AIR 
Comm Stop or HFNC 
Comm Stop alarm in 
standby mode. 
No Comm Stop 
alarm is generated. 
(1) Reconnect or replace the 
cable between the auxiliary 
O2/air or HFNC module 
and the instrument panel. 
A Comm Stop 
alarm is generated. 
(1) Reconnect or replace the 
cable between the auxiliary 
O2/air or HFNC module 
and the motherboard (refer 
to 
11.1.4). 
(2) If the problem persists, 
replace the auxiliary O2/air 
or HFNC module (refer to 
11.1.4). 
(3) If the problem persists, 
2. When the main unit is in 
standby mode, select 
Setup > Service > Data 
Monitors > Power System 
and check whether the 
voltage of the 12VB power 
supply falls within the 
normal range. 
The 12VB power 
supply is abnormal. 
Replace the DC-DC board 
(refer to 11.1.11). 
The 12VB power 
supply is normal. 
 
(1) Reconnect or replace the 
cable between the auxiliary 
O2/air or HFNC module 
and the motherboard (refer 
to 
11.1.4). 
(2) If the problem persists, 
replace the module (refer to 
Segment display 
error of the 
auxiliary O2/air 
or HFNC module 
(no display or 
display error, 
with normal 
backlight of the 
flowmeter) 
/  / 
outlet (refer to 8.1.2) and 
reconnect the cable between 
the auxiliary O2/air or 
HFNC module and the 
instrument panel. 
(2) If the problem persists, 
remove the instrument panel 
(refer to 8.1.5) and 
reconnect the segment 
display to the status screen 
control board. 
(3) If the problem persists, 
replace the segment display 
(refer to 11.1.5). 
(4) If the problem persists, 
replace the auxiliary O2/air