7-44 
MiniMedi CO2 module 
is beyond the normal 
2. If the problem persists, replace the 
CO2 module. 
CO2 Check Cal.  Low 
The MiniMedi CO2 
module fails. 
1. Re-plug the CO2 module. 
2. Recalibrate the CO2 module. 
3. If the problem persists, replace the 
 
7.3.14 NMT Module Runtime Alarms 
Low 
The NMT main cable is 
disconnected. 
Check the connection between the 
main cable and the NMT module. 
NMT No Sensor  Low 
The NMT sensor is 
disconnected. 
1. Check the connection between the 
sensor and the NMT main cable to 
ensure normal connection. 
2. If the alarm persists, replace the 
NMT Stimulation 
Electrode Off 
Low 
The NMT stimulation 
electrode falls off. 
1. Check the connection between the 
stimulation cable and the main cable to 
ensure normal connection. 
2. If the alarm persists, check the 
connection between the stimulation 
electrode and the patient. 
NMT Sensor Comm. 
Error 
Low 
An NMT sensor 
communication error 
occurs. 
1. Check the connection between the 
sensor and the NMT main cable to 
ensure normal connection. 
2. If the alarm persists, replace the 
NMT module or sensor. 
3. If the problem persists, check the 
connection cable between the 
backplane and the display. 
4. If the problem persists, replace the 
module rack (refer to 11.1.14). 
5. If the problem persists, replace the 
cable between the backplane and 
NMT Stimulation 
Current Over Limit 
Low 
The output stimulation 
current of the NMT 
module exceeds the 
limit. 
1. Check the connection between the 
stimulation cable and the main cable to 
ensure normal connection between the 
stimulation electrode and the patient. 
2. If the alarm persists, replace the 
NMT module or related component. 
NMT Comm. 
Abnormal 
High 
NMT communication is 
abnormal. 
1. Re-plug the NMT module, main 
cable, sensor, and stimulation 
electrode, or plug the NMT module to 
another machine. Eliminate the error 
situations one by one. 
2. If the alarm persists, replace the 
NMT module or related component. 
3. If the problem persists, check the