7-1 
7 Troubleshooting 
7.1 Precautions 
Before disassembling the machine for internal testing and maintenance, disconnect the power 
supply and all gas supplies, and shut down the anesthesia machine. 
Before connecting or disconnecting cables or probes inside the machine after disassembly, turn 
off the system power supply to prevent safety risks and damage to the machine. 
7.2 Power-On Troubleshooting 
7.2.1 Power-On and Display Failures 
Cause Confirmation  Confirmation Result  Solution 
When the 
switch is 
turned to 
the ON 
position, 
the main 
screen is 
off. 
AC power supply is 
The AC power supply 
is not connected. 
Connect the AC power supply 
properly. 
2. Check whether the 
AC and battery 
indicators are on. 
Continue checking according to step 
3. 
Continue checking according to step 
5. 
4. Check whether the 
small display 
functions properly. 
The small display is 
off. 
Continue checking according to step 
5. 
The small display 
functions properly. 
Continue checking according to step 
6. 
5. Check whether the 
HFNC module can be 
started. 
The HFNC module 
can be started. 
(1) Replace the system switch (refer 
to 11.1.3). 
(2) If the problem persists, replace 
the DC-DC power board (refer to 
11.1.11). 
(3) If the problem persists, replace 
the backplane (refer to 11.1.11). 
The HFNC module 
cannot be started. 
(1) Replace the DC-DC board (refer 
to 11.1.11). 
(2) If the problem persists, replace 
the backplane (refer to 11.1.11). 
6. Observe whether 
the alarm light blinks, 
whether alarm sound 
is generated, and 
whether the buzzer is 
ringing. 
The alarm light does 
not blink, and no 
alarm sound is 
generated, but the 
buzzer is ringing. 
(1) Remove the rear housing of the 
display by referring to 8.1.7, open 
the service door by referring to 
8.1.1, and reconnect or replace the 
cable between the backplane and the 
display. 
(2) Replace the main board (refer to 
The alarm light blinks, 
and alarm sound is 
generated. 
(1) Remove the rear housing of the 
display by referring to 8.1.7, and 
reconnect the LCD cable or replace 
the LCD assembly (refer to 11.1.1). 
(2) If the problem persists, replace 
the CPU board (refer to 11.1.1). 
Reconnect the display 
cable and LCD cable 
Replace the display (refer to 11.1.1).