7-15 
time and then restart it. If the problem 
persists after the machine runs for a period 
of time (at least 30 minutes), replace the 
DC-DC power board (refer to 11.1.11). 
Heating Module 
Failure 
Low 
1. The temperature 
difference between 
two probes exceeds 
15ºC. 
2. The temperature 
of any heater strip 
exceeds 110ºC for 
consecutive 5s. 
2. If the problem persists, select Setup > 
Service > Data Monitors > Power System 
and check whether the heating module 
temperature and voltage are within the 
normal ranges (based on the ranges of the 
A/D channel). If not, replace the DC-DC 
power board.   
3. During the check in step 2, if the heating 
module temperature is invalid, reconnect 
the cable of the heating module ("huilu" 
socket in the upper left corner of the 
backplane). 
4. If the problem persists, replace the 
heating module (refer to 11.5.3). 
5. If the problem persists, replace the 
DC-DC power board (refer to 11.1.11). 
6. If the problem persists, upgrade the 
software (refer to 11.1.9). 
 
7.3.4 Electronic Flowmeter and BFCS Runtime Alarms 
Electronic Flow 
Control Error 
Med 
CPU voltage error 
occurs. 
2. The 3-way valve in 
the air or N2O branch is 
faulty. 
3. The O2 or balance 
gas branch flow sensor 
or the total flow sensor 
is faulty. 
4. The O2 or balance 
gas branch flow is not 
achieved. 
5. The flow sensor is 
subject to a single 
self-test error. 
6. The two-way valve in 
the EFCS O2 branch is 
2. If the power supply voltage error 
alarm is generated, clear this alarm 
(refer to 7.3.3). 
3. If the problem persists, select 
Setup > Service > Data Monitors > 
FCS to access the flowmeter data 
monitoring screen. Check the voltage 
and sensor flow of the A/D channel. 
4. If the power supply voltage is out 
of the normal range, replace the 
EFCS control board (refer to 11.3.1). 
5. If the measured value of the flow 
sensor is "---" or beyond the normal 
range, replace the flow sensor (refer 
to 11.3.1). 
6. If no fault is detected in steps 4 and 
5, replace the EFCS assembly (refer 
No Fresh Gas  Med 
Fresh gas is turned off 
in manual or 
mechanical ventilation 
mode. 
1. Turn on the fresh gas and set it to a 
proper value. 
2. Check whether gas is supplied 
normally. 
3. If the problem persists, upgrade the 
software (refer to Chapter 6).