7-2 
Cause Confirmation  Confirmation Result  Solution 
powered 
on, the 
main screen 
is on but 
displays no 
of the 
anesthesia 
machine is 
on and can 
display 
content, but 
Reconnect the LCD 
cable (refer to 3.2.2). 
The problem persists. 
(1) Replace the DC-DC power 
board (refer to 11.1.11). 
(2) If the problem persists, replace 
the AC-DC power board (refer to 
11.1.11). 
(3) If the problem persists, replace 
the CPU board (refer to 11.1.1). 
The 
primary 
screen 
functions 
properly, 
but the 
status 
screen 
displays no 
content. 
Check whether the 
Status Screen Comm 
Stop alarm is 
generated. 
The Status Screen 
Comm Stop alarm is 
generated. 
(1) Disassemble the auxiliary outlet 
by referring to 8.1.2 and reconnect 
the cable of the instrument panel. 
(2) If the problem persists, replace 
the status screen control board (refer 
No Status Screen 
Comm Stop alarm is 
generated. 
(1) Disassemble the instrument 
panel by referring to 8.1.5 and 
reconnect the LCD cable in the 
instrument panel (refer to 11.1.5). 
(2) If the problem persists, replace 
the status screen LCD (refer to 
11.1.5). 
(3) If the problem persists, replace 
the status screen control board (refer 
 
7.2.2 Auxiliary Outlet Failures 
auxiliary 
outlet has 
no voltage 
breakers (branch and 
total breakers) of the 
auxiliary outlet are 
Connect the device beyond the 
nominal current to the external 
power supply and reset the breaker. 
 
7.2.3 Leak Test Failures in Auto Mode 
1.  Check the following items, take countermeasures, and perform the leak test again until the 
problem is solved: 
a) Check the pipeline connection: 
  Whether the breathing circuit is connected properly; 
  Whether the leak test plug is connected properly; 
  Whether the sample gas port is sealed or the sampling line is connected properly;